The Phenix line of 5 kV-15 kV AC Hipots offer a completely integrated, easy-to-use high voltage AC supply. The air insulated high voltage transformer, controls and regulator are housed either in a painted steel benchtop enclosure or for the larger units in a free-standing steel enclosure.
All units are equipped with standard safety features such as circuit breaker protection, adjustable output overloads, external interlock provisions, and zero start interlock. The digital meters display a direct reading of the output voltage and current. The output voltage is completely adjustable from near zero to 100% of output via a rugged and reliable variable autotransformer.
Model | 605-2P | 605-5P | 605-10P | 610-2P | |
Input |
Voltage | 115/120 V, 60 Hz or 220/240 V, 50 Hz |
220/240 V | 220/240 V | 115/120 V, 60 Hz or 220/240 V, 50 Hz |
Frequency | 115/120 V, 60 Hz or 220/240 V, 50 Hz |
50/60 Hz | 50/60 Hz | 115/120 V, 60 Hz or 220/240 V, 50 Hz |
|
Output |
Voltage | ≈ 0-5 kV | ≈ 0-5 kV | ≈ 0-5 kV | ≈ 0-10 kV |
Current | 0-400 mA | 0-1 A | 0-2 A | 0-200 mA | |
Duty Cycles |
5 min ON/15 min OFF | 2 kVA | 5 kVA | 10 kVA | 2 kVA |
Option Burn |
Rating | 1 kVA | 1 kVA | 3 kVA | 1 kVA |
Model | 610-10P | 610-20P | 615-10P | 615-20P | |
Input |
Voltage | 220/240 V | 220/240 V | 220/240 V | 220/240 V |
Frequency | 50/60 Hz | 50/60 Hz | 50/60 Hz | 50/60 Hz | |
Output |
Voltage | ≈ 0-10 kV | ≈ 0-10 kV | ≈ 0-15 kV | ≈ 0-15 kV |
Current | 0-1 A | 0-2 A | 0-667 mA | 0-1.33 A | |
Duty Cycles |
5 min ON/15 min OFF | 10 kVA | 20 kVA | 10 kVA | 20 kVA |
Option Burn |
Rating | 3 kVA | 3 kVA | 3 kVA | 3 kVA |
safety and design features
- One-piece design
- Circuit breaker protection
- External interlock provision
- External warning circuit provision
- Latching HV On/Off pushbuttons
- Zero Start Interlock
- Manual control of output voltage
- Adjustable overload circuit
- Slow and fast-acting transient protection on all meters and relays
- Digital Meters (3 1/2 digit LCD display)
- Digital Timer with audible alarm
- Emergency Off pushbutton
- Output overload indicator with reset switch
- Casters mounted on cabinets (excludes -2P models)
- Input power cable 5’ (1.5 m) and shielded output cable 15’ (4.5 m) are included
- Operation/maintenance manual