The Tektronix PA1000 is a single-phase, single-channel power analysis solution that is optimized for fast, efficient, and accurate power consumption testing to international standards. It’s compact size, DMM-like user-interface, graphical display, and powerful software enable users to quickly visualize, analyze, and document the power consumption efficiency of next-generation devices, including standby power measurements and harmonic analysis.
- Voltage and current ranges
- Measurement accuracy - voltage
-
- Voltage accuracy, VRMS (45 Hz to 850 Hz)
- ± 0.04% of Reading ± 0.04% of Range ± 0.005 V
- Voltage accuracy, VRMS (10 Hz to 45 Hz, 850 Hz to 1 MHz, typical)
- ± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± 0.05 V
- Voltage accuracy, DC (typical)
- ± 0.1% of Reading ± 0.1% of Range ± 0.05 V
- Effect of common mode (typical)
- 100 V, 100 kHz
- Measurement accuracy - current
-
- Current accuracy, ARMS (45 Hz to 850 Hz) 1
- ± 0.04% of Reading ± 0.04% of Range ± (1.8 µV/Zext)
- Current accuracy, ARMS (10 Hz to 45 Hz, 850 Hz to 1 MHz, typical)
- ± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± (50 µV/Zext)
- Current accuracy, DC (typical)
- ± 0.1% of Reading ± 0.1% of Range ± (100 µV/Zext)
- Current - peak inrush accuracy (100 Apeak range)
- 2% of Range ± 20 mA
- Effect of common mode (typical)
- 100 V, 100 kHz, 20 A shunt 100 V, 100 kHz, 1 A shunt 100 V, 100 kHz, external shunt
| Dimensions | mm | in |
|---|---|---|
| Height | 102 | 4.0 |
| Width | 223 | 8.7 |
| Depth | 285 | 11.2 |
| Weight | Kg | lb |
| Net (without lead set) | 3.2 | 7.0 |
| Temperature | C | F |
| Operating | 0 °C to +40 °C | +32 °F to +102 °F |
| Nonoperating | -20 °C to +60 °C | -4 °F to +140 °F |