Legendary support and quality meets ease-of-use and affordability. The Tektronix TTR500 series 2-port, 2-path VNA is our latest breakthrough – an unmatched combination of measurement performance and convenience. Get the power to make everyday measurements with the accuracy and confidence you expect from Tektronix, all without breaking your budget.
| Frequency | Directivity (dB) | Source match (dB) | Load match (dB) | Reflection tracking (±dB) | Transmission tracking (±dB) |
|---|---|---|---|---|---|
| 300 kHz to < 500 kHz | -47.5 | -36.6 | -47 | 0.062 | 0.066 |
| 500 kHz to < 2MHz | -50.7 | -39.6 | -47 | 0.062 | 0.042 |
| 2 MHz to < 10 MHz | -50.7 | -39.6 | -47 | 0.062 | 0.042 |
| 10 MHz to < 200 MHz | -50.7 | -39.6 | -47 | 0.062 | 0.042 |
| 200 MHz to < 1.5 GHz | -46.6 | -34.8 | -37.1 | 0.08 | 0.073 |
| 1.5 GHz < 3 GHz | -46.6 | -34.8 | -37.1 | 0.08 | 0.073 |
| 3 GHz to 6 GHz | -39.7 | -25.3 | -36 | 0.166 | 0.156 |
| 18 °C to 28 °C, within 1 °C of calibration temperature and at the same ambient humidity conditions in which calibration was performed. | |||||
Key performance specifications
- 100 kHz to 6 GHz frequency range
- > 122 dB dynamic range
- -50 to +7 dBm output power
- < 0.008 dB RMS trace noise
Key features
- Full 2-port 2-path S-parameter measurement (S11, S21, S12, S22) in a variety of formats.
- Complete vector network analysis capability with Tektronix VectorVu-PC™ software.
- Built-in bias tee accessible on both ports to bias active devices.
- Application programming interface (API) for Microsoft Windows environment and LabView driver.
- Robust SCPI command interface (compatible with current VNA models) to optimize code migration.
- Touchstone file support (import/export) for use with Electronic Design Automation (EDA) simulation tools.
- Offline simulation mode for analysis of S-parameter files when disconnected from the instrument.
- Industry-leading three year warranty.
- Time domain with gating.
- Automatic calibration support.
Comments are closed.